Medtronic Paradigm Link Guida Utente Pagina 211

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 210
#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 70 mg/dL (3.9
mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
= -0.5 units
+
(correction estimate)
= 2 + (-0.5)
= 1.5 units
bolus estimate
3.9 mmol/L - 5.0 mmol/L
2.2 mmol/L/u
= -0.5 units
70 mg/dL - 90 mg/dL
40 mg/dL/u
#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 100 mg/dL (5.6
mmol/L).
(food estimate)
60 g
30 g/u
= 2 units
+
(correction estimate)
= 2 + 0
= 2 units
bolus estimate
Correction is 0 because the
current blood glucose reading
is between the BG High and
Low Targets
Pump specifications
192
Vedere la pagina 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 261 262

Commenti su questo manuale

Nessun commento