Vedere la pagina 232
The percent of Guardian RT readings in the previous graph are presented in the following table according
to the percentage of points falling within each zone (A-E). Results are further broken down (stratified)
according to the range of glucose concentrations.
Glucose
Range
(mg/dl)
Number and
(%) of Data
Points
Evaluated
A + B A B C D E
40-80 356 (9) 271 (76.1) 214 (60.1) 57 (16.0) 2 (0.6) 80 (22.5) 3 (0.8)
81-120 769 (20) 768 (99.9) 463 (60.2) 305 (39.7) 1 (0.1)
N/A
a
N/A
121-240 2362 (60) 2352 (99.6) 1476 (62.5) 876 (37.1) 4 (0.2) N/A 6 (0.2)
>240 454 (11) 394 (86.8) 277 (61.0) 117 (25.8) N/A 59 (13.0) 1 (0.2)
Overall 3941 (100) 3785 (96.0) 2430 (61.7) 1355 (34.4) 7 (0.2) 139 (3.5) 10 (0.2)
a. N/A means that the Clarke Error Grid does not consider the possibility of these zones in that concentration range.