Medtronic Paradigm 522 Guida Utente Pagina 231

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 230
The percent of Guardian RT readings in the previous graph are presented in the following table according
to the percentage of points falling within each zone (A-E). R
esults are further broken down (stratified)
according to the range of glucose concentrations.
Glucose
Range
(mg/dl)
Number and
(%) of Data
Points
Evaluated
A + B A B C D E
40-80 356 (9) 271 (76.1) 214 (60.1) 57 (16.0) 2 (0.6) 80 (22.5) 3 (0.8)
81-120 769 (20) 768 (99.9) 463 (60.2) 305 (39.7) 1 (0.1)
N/A
a
N/A
121-240 2362 (60) 2352 (99.6) 1476 (62.5) 876 (37.1) 4 (0.2) N/A 6 (0.2)
>240 454 (11) 394 (86.8) 277 (61.0) 117 (25.8) N/A 59 (13.0) 1 (0.2)
Overall 3941 (100) 3785 (96.0) 2430 (61.7) 1355 (34.4) 7 (0.2) 139 (3.5) 10 (0.2)
a. N/A means that the Clarke Error Grid does not consider the possibility of these zones in that concentration range.
Appendix A212
Vedere la pagina 230
1 2 ... 226 227 228 229 230 231 232 233 234 235 236 ... 261 262

Commenti su questo manuale

Nessun commento