Medtronic Guardian Real-time Guida Utente Pagina 119

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 118
5 Prepare a mild liquid soap solution using one (1) teaspoon of Ivory liquid soap per one (1) gallon of
room temperature tap water.
6 With tester still attached, submerge the transmitter in the mild liquid soap solution and soak for one
minute.
7 Holding the tester, brush the entire surface of the transmitter using a soft-bristled toddler toothbrush.
Make sure to brush all hard-to-reach areas until visibly clean.
8 Rinse the transmitter under running room temperature tap water for at least one minute, and until all
visible liquid soap is gone.
9 Dry the transmitter and tester with a clean, dry cloth.