Time of day
Glucose Conc (mg/dL)
The above chart, Guardian RT device #2 is represented by a solid line, and the Guardian RT device #4 by a
light-gray line.
Low and High Alerts in adults
The ability of the Guardian RT to detect high and low glucose levels was measured in the same clinical
study.
6
Since it is important to set the alert levels in a conservative fashion, the Low Glucose Alert should
be set at a value slightly higher than the value of blood glucose you want to detect, and the High Glucose
Alert should be set at a value slightly lower than the value of blood glucose you want to detect.
NOTE: Please ask your healthcare professional which low and high alert setting is best for you.
6. Medtronic Diabetes, A Frequent Sample Accuracy Evaluation of the Medtronic Diabetes Telemetered Glucose Monitoring
System II (TGMS II) in Subjects with Type 1 Diabetes Mellitus, August 2004.
Sensor accuracy 139
Appendix A